Current research projects

Image Tribological investigations of oil-refrigerant-material-systems
Image Thermostatic Expansion Valves
Image Thermal engines
Image High Capacity Pulse Tube Cooler
Image Cryostats, Non-Metallic and Metallic
Image Influenced melting point of water by magnetic field
Image Brine (water)-water heat pump
Image Ionocaloric cooling
Image Mass Spectrometer
Image Lifetime prediction of hermetic compressor systems
Image Certifiable connection types in cryogenics
Image Investigation of material-dependent parameters
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Software for test rigs
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Hydrogen and methane testing field at the ILK

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary