Current research projects

Image Cryostats, Non-Metallic and Metallic
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Verification of storage suitability of cryo tubes
Image 3D - Air flow sensor
Image Humidifier System for High-Purity Gases
Image Computational fluid dynamics CFD
Image All-in-one device for freeze-drying and production of biomaterial
Image Thermostatic Expansion Valves
Image Air-water heat pumps
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Reduction of primary noise sources of fans
Image Corrosion inhibitor for ammonia absorption systems
Image Non- invasive flow measurements
Image Combined building and system simulation
Image Test procedures for electrical components

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary