Current research projects

Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Heat2Power
Image State of system and failure analyses
Image Influenced melting point of water by magnetic field
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Helium extraction from natural gas
Image Practical training, diploma, master, bachelor
Image Testzentrum PLWP at ILK Dresden
Image Panel with indirect evaporative cooling via membrane
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Laseroptical measurement
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Verification of storage suitability of cryo tubes
Image Reduction of primary noise sources of fans
Image Modular storage system for solar cooling
Image Filter Tests

You are here:   /  Home


Low noise and non metallic liquid-helium cryostat

Industry and Research Institutes

Dr. rer. nat. Andreas Kade

+49-351-4081-5117

Low-noise Magnetic Field Cryostat for SQUID-Applications

Technical Data for the Cryostats

parametervalue
volume of liquid helium5 to 12 litre (other volume possible)
time to complete evaporation of helium3 to 7 days (depends on size)
heat load0.1 W (in standby operation)
rate of helium evaporation≤ 3 litre / day
holding time of the liquid heliumup to 4 days without refilling
initial noise of the cryostat< 3 fT / Hz½
helium leak test (He-vessel)< 1 × 10-11 mbar l / s

ILK Dresden developed low noise magnetic field and non metallic helium cryostats made of GRP (glass reinforced plastics) with a high helium and vacuum holding time.

The cryostats can operate in a direction independent operation mode (tiltable arrangement) and hence they are suitable for a plurality of several applications.

The ILK concept offers much lower permeation rates inside the He-reservoir as comparable models from other manufacturers and is therefore perfect for SQUID (superconducting quantum interference device) applications and other long-term measurements.
 
The technical design of the cryostats is thus predestined for future sensor generation.

Specification

  • suitable for the cooling of SQUID-Sensors
  • non-metallic (GRP)
  • fast filling with LHe via thermosiphon
  • GRP exhaust gas cooled radiation shield for high efficiency.
  • small cold-warm distance feasible
  • low maintenance 
  • other design variants possible on costumer request

Low noise magnetic field

The low-noise magnetic field of the cryostat was tested in the magnetically shielded room BMSR-1 of PTB Berlin and is smaller than the lowest resolution limit of the measurement system used, see figure below.