Current research projects

Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Test rigs for refrigeration and heat pump technology
Image Low noise and non metallic liquid-helium cryostat
Image Reduction of primary noise sources of fans
Image Testing of mobile leak detectors according to DIN EN 14624
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Pulse-Tube-Refrigerator with sealed compressor
Image Combined building and system simulation
Image Electrical components in refrigeration circuits
Image Low Temperature Measuring Service
Image Refrigerants, lubricants and mixtures
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Performance tests of condensing units
Image Testzentrum PLWP at ILK Dresden
Image Overall System Optimization of Refrigeration Plant Systems for Energy Transition and Climate Protection
Image Cold meter

You are here:   /  Home


Mass Spectrometer

Industrie und Forschungsinstitute

Gunar Schroeder

+49-351-4081-5129

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Mass spectrometry is a method for determining the composition of gas mixtures in the high or ultra-high vacuum range. It can be used to carry out qualitative and quantitative gas analyses and leak detection or to determine trace contamination.

The individual neutral gas particles are ionized by an electron impact ion source and separated from each other in an electric quadrupole field according to their mass-to-charge ratio. The filtered ions are then recorded by a detector. The ion current detected is proportional to the partial pressure of the respective gas component. For a more precise quantitative gas analysis, the relatively operating measuring instrument can be calibrated with a calibration gas.

The test set-up consists of a test chamber with a connected turbo-molecular-pump and fore vacuum pump. The mass spectrometer requires a vacuum <1,5×10-7 psi (<1×10-5 mbar) to be able to measure with high precision. The mass spectrometer and a vacuum sensor are mounted on the test chamber. The mass spectrometer is connected to a measuring PC with corresponding software.

Available measurement modes are:

  • Vacuum analysis, to determine residual gas components, evaluation as a plot of ion current and mass
  • Leak detection, with helium or another tracer gas, the ion current of the tracer gas or the corresponding mass is recorded and plotted over the time
  • Mass spectrum, evaluation of ion current and mass with user-defined specifications and time trend
Paramter Test value
Analysis Vacuum analysis, leak detection, mass spectrum, trace contamination
Detectable gases H2, He, N2, H2O, O2, CO2 usw.
Mass range 1 – 100 Da (1,66×10-27 – 1,66×10-25 kg)
Detection limit 1.5×10-15 psi / 3×10-13 mbar
Min. vacuum pressure and max. temperature range <7.3×10-6 psi / 5×10-4 mbar, < 150°C / 302°F

[Translate to EN:]


Your Request

Further Projects