Current research projects

Image High Capacity Pulse Tube Cooler
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Cool Up
Image Low Temperature Measuring Service
Image Innovative Parahydrogen Generator Based on Magnets
Image Thermal engines
Image Refrigerants, lubricants and mixtures
Image Service offer for Leak Detection and Tightness Test
Image Energy efficiency consulting - cogeneration systems
Image Computational fluid dynamics CFD
Image Low noise and non metallic liquid-helium cryostat
Image Swirl-free on the move...
Image Software for technical building equipment
Image Measurements on ceiling mounted cooling systems
Image Measurement of insulated packaging
Image Influenced melting point of water by magnetic field

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary