Current research projects

Image Multifunctional electronic modules for cryogenic applications
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Air-water heat pumps
Image 3D - Air flow sensor
Image Tribological investigations of oil-refrigerant-material-systems
Image Cryostats, Non-Metallic and Metallic
Image Test procedures for electrical components
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Cool Up
Image Humidifier System for High-Purity Gases
Image Measurements on ceiling mounted cooling systems
Image Service offer for Leak Detection and Tightness Test
Image Cold meter
Image State of system and failure analyses
Image Electrical components in refrigeration circuits

You are here:  Home /  Testing Equipment


Humidifier System for High-Purity Gases

Industry and Research Instituts

Moritz Kuhn

+49-351-4081-5120

Nafion - moisture transfer

At the ILK Dresden a humidification system was developed and manufactured which is especially suitable for research facilities where very low dew point temperatures have to be adjusted and at the same time a high gas purity has to be guaranteed. The core component is a PermaPure® humidifier with Nafion® moisture exchange. The humidification system precisely controls the dew point via a bypass mixing with dry gas and also takes over the automatic filling of evacuated containers. The control system is optimised to prevent recondensation. Humidification can be controlled from any plant system via an interface. Table 1 shows the technical data of a humidification system for the humidification of synthetic air realized for the KIT in Karlsruhe. The system can be adapted for other applications, and all parameters can be adjusted to the respective requirements.

Table  1: Technical Data Humidifier System
Parameter Value
setpoint gas flow 750 litre/min
maximum relative humidity 95%
minimum dew point -60°C / -76°F
degree of gas purity 5.0
gas purity, number of particle < 5 particle/cm³