Current research projects

Image Hybrid- Fluid for CO2-Sublimation Cycle
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Behavior of multiphase cryogenic fluids
Image Service offer for Leak Detection and Tightness Test
Image Investigation of material-dependent parameters
Image Brine (water)-water heat pump
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Thermal engines
Image Low Temperature Tribology
Image Investigation according to DIN EN ISO 14903
Image Measurement of insulated packaging
Image Micro heat exchangers in refrigeration
Image Calibration leak for the water bath leak test
Image Corrosion inhibitor for ammonia absorption systems
Image Influenced melting point of water by magnetic field
Image Swirl-free on the move...

You are here:  Home /  Testing Equipment


Low noise and non metallic liquid-helium cryostat

Industry and Research Institutes

Dr. rer. nat. Andreas Kade

+49-351-4081-5117

Low-noise Magnetic Field Cryostat for SQUID-Applications

Technical Data for the Cryostats

parametervalue
volume of liquid helium5 to 12 litre (other volume possible)
time to complete evaporation of helium3 to 7 days (depends on size)
heat load0.1 W (in standby operation)
rate of helium evaporation≤ 3 litre / day
holding time of the liquid heliumup to 4 days without refilling
initial noise of the cryostat< 3 fT / Hz½
helium leak test (He-vessel)< 1 × 10-11 mbar l / s

ILK Dresden developed low noise magnetic field and non metallic helium cryostats made of GRP (glass reinforced plastics) with a high helium and vacuum holding time.

The cryostats can operate in a direction independent operation mode (tiltable arrangement) and hence they are suitable for a plurality of several applications.

The ILK concept offers much lower permeation rates inside the He-reservoir as comparable models from other manufacturers and is therefore perfect for SQUID (superconducting quantum interference device) applications and other long-term measurements.
 
The technical design of the cryostats is thus predestined for future sensor generation.

Specification

  • suitable for the cooling of SQUID-Sensors
  • non-metallic (GRP)
  • fast filling with LHe via thermosiphon
  • GRP exhaust gas cooled radiation shield for high efficiency.
  • small cold-warm distance feasible
  • low maintenance 
  • other design variants possible on costumer request

Low noise magnetic field

The low-noise magnetic field of the cryostat was tested in the magnetically shielded room BMSR-1 of PTB Berlin and is smaller than the lowest resolution limit of the measurement system used, see figure below.