Current research projects

Image Cold meter
Image Range of services laboratory analyses
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Intelligent innovative power supply for superconducting coils
Image Panel with indirect evaporative cooling via membrane
Image Filter Tests
Image Mass Spectrometer
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Tribological investigations of oil-refrigerant-material-systems
Image IO-Scan - Integral measuring optical scanning method
Image Modular storage system for solar cooling
Image Investigation of materials
Image Innovative small helium liquefier
Image Investigation according to DIN EN ISO 14903
Image Pulse-Tube-Refrigerator with sealed compressor
Image State of system and failure analyses

You are here:   /  Home


Calibration of Low Temperature Sensors

141,142

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary